Émetteurs récepteurs

Préamplificateur d'émetteur à distance 5627 RPP-1

The performance of high frequency ultrasonic transducers can be significantly affected by the length of the cable between the transducer and pulser-receiver. The Panametrics-NDT Model 5627RPP-1 Remote Pulser Preamplifier permits use of an optimum short cable in applications when setup requirements would otherwise dictate a cable length that would degrade transducer performance.

Features & Benefits

•  High Frequency broadband receiver (10-150 MHz)
•  24 dB gain
•  Low noise
•  Switchable energy control
•  Pulser repetition rate 0 to 20 KHz
•  Rise Time < 2nS
•  Compatible with Model 5900PR Pulser-Receiver
•  Able to drive 500 ft. (152 m) cable

Applications

The Model 5627RPP-1 is designed for imaging, flaw detection, thickness gaging, and material characterization applications in the frequency range of 20 to 125 MHz. It is used with a host instrument, normally a Panametrics-NDT Model 5900PR Pulser-Receiver. Receiving power and an external trigger command from the host instrument, the Model 5627RPP-1 generates an excitation pulse of selectable energy. The excitation pulse is transmitted to the transducer through a very short (typically 12 inch/305 mm) coaxial cable. Returning echoes are amplified by a high frequency broadband preamplifier, and then sent to the host instrument for further processing. Cable length between the Model 5627RPP-1 and the host instrument can be as great as 500 ft./152 m.
Because the cable connecting the Model 5627RPP-1 to the transducer is very short, undesirable cable effects are minimized. However, the instruments’ ability to drive a very long cable to the host instrument provides great mechanical flexibility in setups. For contact applications it can be easily hand-carried. Mechanical dimensions and an interconnection diagram are on the reverse side of this data sheet. .

Specifications*

PULSER 5627 RPP-1
Repetition Rate:0 to 20 KHz by external trigger pulse
Pulse Type
(main bang):
Negative Impulse
Maximum Pulse150 Volts at minimum damping  (damping Amplitude: control fully ccw)
Damping:

6 to 50 ohms ± 10% continuously  adjustable

Rise Time:<2nS (50 ohms damping)
Available Pulse Energy:1, 2, 4, and 8uJoules, switch selectable
Ext. Trig. Requirements:

TTL COMPATIBLE
Rise Time: <20nS
Pulse Width: 200nS # PW # 1uS
When used with the Model 5900PR  compatible trigger is furnished by the host instrument

RECEIVER 5627 RPP-1
Voltage Gain:
(Non-Inverting)
24 dB ±2 dB
(RL = 50 ohms and 4ft.  RG58/U cable)
Bandwidth (-3dB)10 MHz to 150 MHz.
Noise:Typically 60u V pk-pk referred to the input (BW = 150 MHz), 100uV pk-pk max.
Output Impedance:

50 ohms

Maximum Output Voltage:±0.3 Volt peak (RL= 50 ohms)
Cable Length:10 ft. (3 m) standard cable length 500 ft (152 m) recommended max
Operating Temperature:32° to 122° F (0 to 50° C)
UNIT 5627 RPP-1
Power Requirements:
Pin 5:
Pin 1:
Pin 2:
Pin 3:

+260 V to 300V at 10ma
+12 V ± 5% at 170ma
-12 V ± 5% at 60ma
Ground

Power Input:

HIROSE HR10A-7R-6P
Trigger Input:BNC
Signal Output:BNC
Transducer:BNC
Operating Temperature:0º to 50ºC ambient
Weight:1.3 lbs. (590 grams)