Détail de l'application
Ultrasonic Thickness Measurements of Hot Metals
Ultrasonic thickness gauging is not just limited to materials at normal (ambient) temperatures. The thickness of a
part can be measured even when the surface temperature approaches 900 deg F (500 deg C). In metals, particularly,
there may be circumstances when the thickness of the part needs to be measured before the part can be cooled
down.
Ultrasonics has the advantage - just as at normal temperatures - that thickness measurements can be taken from one
side of the hot part. Ultrasonics may be the preferred method of measurement even when both sides of the hot metal
part are accessible. The reason is that ultrasonic thickness measurements can be made instantly and with very brief
contact with the hot surface.
For the best calibrated accuracy, the effect of the higher temperature on the velocity of sound must be taken into
account. Although change in sound velocity with temperature is much less in metals than in plastics, significant
errors in measured thickness can result if sound velocity is not corrected.
Equipment: A major reason to classify hot metals separately is that when the temperature of a
surface exceeds 120 deg F (50 deg C), Microscan contact transducers cannot be used because they will
be damaged by thermal expansion effects. Consequently, gauges that exclusively utilize contact transducers cannot be
used for making thickness measurements on hot surfaces.
Instead, we recommends the Model 35DL, which can be used with delay line
transducers or immersion transducers. On hot surfaces delay line transducers must be used
with High Temperature delay lines.
The thickness of hot corroded metal parts or parts with rough surfaces should be measured with the Model 37DL
PLUS and a D790 or D791 dual transducer. The 37DL PLUS is a sophisticated corrosion gauges
with internal data loggers and waveform displays that have practical features to provide excellent thickness
measurement reliability on hot corroded metal pipes or tanks. For less challenging applications, the compact
MG2 series of gages can also be used. |
|
|