Industrial Testing Instruments

µTomoscan Portable Ultrasound Inspection System

Obsolete Products

This product has been replaced by the OmniScan MX

The µTomoscan is our most economical multichannel UT system. It allows the use of pulse-echo, transmission technique and time-of-flight-diffraction technique (TOFD). It features both logarithmic and linear amplifiers. The µTomoscan is ideal for in-service inspection that requires automated data acquisition, real-time imaging, flaw sizing, etc.The µTomoscan system is built around R/D Tech Tomoview™ a standardized Windows-based software platform for data acquisition and analysis. This software offers the flexibility to configure layouts for any specific application to better meet your particular inspection requirements.

General Features

  • Up to 8 internal ultrasound channels expandable to 16 remote ultrasound channels
  • Pulse echo (P/E), time-of-flight diffraction (TOFD), or transmission mode on any channel
  • 20 kHz pulsing rate (PRF)
  • 100 MHz, 8-bit A/D converter
  • DAC curves
  • Real-time averaging of waveforms
  • Real-time data compression
  • Real-time waveform recording
  • Two measurement gates and one synchronization gate per channel
  • Logarithmic and linear amplifiers
  • Real-time display of A-scan, B-scan, C-scan, and D-scan
  • Echodynamic and envelope displays
  • Multichannel, angle-corrected real-time display of top, side, and end views
  • Merged imaging of data from multiple channels
  • Flaw representation on 2D overlays
    (using imported 2D and 3D CAD files)
  • Direct interface to manual and motorized scanners
  • Multitasking system:
    Analysis may proceed during data acquisition
  • PC-based setup, data acquisition, and data analysis
  • Storage of unlimited number of user-defined inspection setups
  • Reports using user-defined layouts

Weld inspection
Based on the Olympus NDT's strong know-how in nuclear, petrochemical, and heavy industry, the µTomoscan includes all the functions required to perform weld inspections. The multichannel µTomoscan offers the possibility to combine TOFD and P/E techniques, which allows more complete and accurate characterization and sizing of defects.
Corrosion mapping and surface inspection
In petrochemical applications, corrosion detection and mapping represent a major portion of the inspection requirement. The aerospace market also requires scanning of large surfaces for disbonding and delamination.

Heavy industry
The use of automated ultrasound NDT equipment in heavy industry provides immediate inspection results and superior flaw sizing. Compared to manual inspections, automated inspection systems guarantee complete coverage, and provide supporting documentation to that effect. Inspection results are provided directly to a computer, and data may be archived and printed.

Industrial NDT
The open architecture design of the hardware and software of the µTomoscan enables it to adapt to plant operation and go-no-go process control. The µTomoscan packaged in an industrial housing along with an industrial computer and MCDU-02 in rack-mount version meet these requirements.

Specifications for the µTomoscan

Physical Specifications
Internal channels8 channels
Encoder interfaceQuadrature 2-encoder input
CommunicationsStandard Ethernet™
ExpandabilityRemote pulser/receiver up to 16 channels
Motor controlCompatible with the MCDU-02
Data Acquisition
Digitizer8 bits, up to 100 MHz
(50, 25, 12.5, 6.25, 3.125 MHz)
A-scan length32 to 8,096 points
Acquisition delay0 ms to 2 ms
Acquisition modeFree running, encoder position, or external signal
SynchronizationPulser, surface echo, external trigger
Real-time averaging1, 4, 8, and 16
Data Acquisition Performance
A-scan recording1,200 waveforms/s (512 samples)
Peak data5,000 data points per second
C-scan data20,000 data points per second (A1, A2, T1, and T2)
Data transfer600 KBps
Pulser/receiver
Pulse width25 ns to 1000 ns
Pulser amplitudeUp to 300 V unipolar negative pulse
System bandwidth0.5 MHz to 24 MHz (-3 dB)
Linear amplifier gain0 dB to 100 dB, 0.5-dB steps
High-pass filterNone, 1, 2, 5, and 10 MHz
Low-pass filterNone, 2, 5, 10, and 15 MHz
Input noise-83 dB
Logarithmic amplifierSigned with 80 dB dynamic range
Data Type
File formatRDTIFF
A-scanRF, rectified, or rectified video
Peak dataMaximum of 128 peaks per waveform
C-scan typeGate 1 and 2 (amplitude and time of flight)
Power and Environment
Voltage and frequency90/132 V 60 Hz, 180/264 V 50 Hz
DAC
Dynamic range80 dB
Slope20 dB/µs, 30 dB/µs for logarithmic amplifier
Manual calibration
Automatic calibration from envelope mode
Ramp
Alarm and I/O
Up to 8 analog inputs
External trigger
External reset
External start
2 analog outputs for amplitude of time of flight in gates
Digital output for multiplexing
3 alarm outputs coming from any combination of 2 gates
English | 日本語 | français | 中文 | Deutsch | italiano | čeština | magyar Español | Tiểng Việt |