Thiết bị đo chiều dày cầm tay
|
The Panametrics- NDT™ 25MX PLUS is an ultrasonic precision thickness gage that provides cost-effective solutions in applications where the opposite side of the test material is difficult or impossible to reach. Multipoint and single thickness measurements can be made on a wide variety of materials including metal, plastic, rubber, glass, fiberglass, and composites.
|
Features
|
|
• Single or multipoint thickness measurements
• Connects up to 8 transducers
• User selectable: number of channels, channel sequencing, and channel rate
• Uses contact, delay line, and immersion transducers
• Transducer frequency range 2 - 30 MHz
• Wide thickness range, from 0.003 to 20 inches (0.080 mm - 500 mm) depending on application
• Resolution up to 0.0001" (0.001 mm)
• A-scan display for waveform verification
• Application Auto-Recall with stored standard and custom setups
• Alphanumeric file-based datalogger stores 18,000 thickness readings or 1,750 waveforms
• Handheld, intuitive operation, long battery life
|
MX-8 Multiplexer
|
|
Number of Channels: 1 to 8, user selectable
Channel Switching Rate: 1, 2, 4, 8, 16, or 20 Hz, user selectable
Size: 3.2" x 5.9" x 1.25"(149.9 mm x 81.3 mm x 31.8 mm)
Weight: 7.7oz. (.22 kg)
Multipoint Thickness Measurements Connected to the optional MX-8 multiplexer, the 25MX PLUS has the ability to collect and display thickness measurements from up to 8 separate transducers.
Special software on the 25MX PLUS
allows the user to select:
• Number of channels (1-8)
• Mux Mode
- Off
- Single Channel (manual switching)
- Single Sequence (switch through selected channels on command)
- Sequence (continuous channel switching)
• Channel Rate (1, 2, 4, 8, 16 or 20 Hz; thickness output and channel number)
Single Thickness Measurements
The 25MX PLUS operates as a 'standard' precision thickness gage when not connected to the MX-8 Multiplexer.
|
25MX PLUS Specifications*
|
Measurements| Mode 1: | Time interval between excitation pulse and first backwall echo, using contact transducers. | | Mode 2: | Time interval between the first interface echo after the excitation pulse and the first backwall echo, using delay line or immersion transducers. | | Mode 3: | Time interval between successive backwall echoes following the first interface echo after the excitation pulse, using delay line or immersion transducers. |
Thickness Measurement Range: Steel: 0.006 - 20.000 inches
(0.150 - 508.00 mm)
Plastic: 0.003 - 2.0000 inches
(0.080 -50.00 mm)
Thickness range depends on material, transducer, surface condition, temperature, and set up selected.
Material Velocity Range: 0.02000 - 0.55110 inch/µS
(0.5080 - 13.9979 mm/µS)
Resolution: Keypad Selectable: | LOW: | 0.01" | 0.1 mm | | STANDARD: | 0.001" | 0.01 mm | | HIGH: | 0.0001" | 0.001 mm |
Battery: 6 V Rechargeable NiCad battery pack, or field-replaceable alkaline AA batteries
Battery Life: 25 hours in normal measurement mod | |